Assessment of error in the MV radiation isocenter position calculated with the Elekta XVI software
Autor: | Jonathan R Sykes, George B. Warr, David Thwaites, Jacek M. Chojnowski |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2020 |
Předmět: |
Rotation
Elekta XVI software quality assurance Software error Radiation Linear particle accelerator Coincidence 030218 nuclear medicine & medical imaging 03 medical and health sciences 0302 clinical medicine Optics Software Technical Note Humans Radiology Nuclear Medicine and imaging Instrumentation radiation isocenter Physics business.industry Phantoms Imaging Rounding Isocenter ball‐bearing 87.55Qr 030220 oncology & carcinogenesis Technical Notes Particle Accelerators business Quality assurance EPID Algorithms |
Zdroj: | Journal of Applied Clinical Medical Physics |
ISSN: | 1526-9914 |
Popis: | The assessment of the coincidence of imaging and radiation isocenters is an important task of regular quality assurance of medical linear accelerators (linacs) as recommended in national and international quality assurance guidelines. A previously reported investigation of the accuracy of the Elekta XVI software to localize the linac radiation isocenter, by comparing statistically with other independent software, has shown some discrepancies at the sub‐mm level. A further investigation is carried out here using a set of reference images and mathematical operations to observe how the Elekta XVI software analyses them. Symmetric mathematical operations on reference images should result in symmetrical outcomes. Three different rotation functions are used in increasing degree of complexity to characterize the Elekta XVI software error in the linac radiation isocenter position. No independent algorithms or phantoms are used in this methodology. The magnitude and direction of the radiation isocenter localization error has been determined to be consistently 0.13 mm or 0.14 mm in the longitudinal direction towards the target depending on the case. The radiation isocenter localization error comprises two separated errors of the Ball Bearing Center by 0.13 mm and MV Field Center by either 0.00 mm or −0.01 mm in the longitudinal direction towards the target. The calculation of the MV Field Center is influenced by the polymethyl methacrylate rod supporting the ball‐bearing. The precise value and the root cause of the error cannot be assessed due to the rounding effect of the results reported by the Elekta XVI software and lack of access to the source code. |
Databáze: | OpenAIRE |
Externí odkaz: |