Single-center experience with the Tigertriever device for the recanalization of large vessel occlusions in acute ischemic stroke

Autor: Omer Yildiz, Bedih Balkan, Hatem Hakan Selcuk, Hasanagha Zalov, Batuhan Kara, Aysun Erbahceci Salik, Günay Gül
Rok vydání: 2018
Předmět:
Zdroj: Journal of neurointerventional surgery. 11(5)
ISSN: 1759-8486
Popis: BackgroundTigertriever (Rapid Medical, Yoqneam, Israel) is a new design of stent retriever.PurposeTo evaluate the feasibility, safety, and efficacy of the Tigertriever in patients with acute ischemic stroke who have undergone mechanical thrombectomy.Materials and methodsTwo different techniques—namely, standard unsheathing (SUT) and repetitive inflation-deflation (RID) techniques, were used. Modified Thrombolysis in Cerebral Infarction (mTICI) scores of 2b and 3 were considered as successful recanalization.ResultsA total of 61 thrombectomy procedures with Tigertriever were retrospectively evaluated. The mean age of patients was 60.7 years and their National Institutes of Health Stroke Scale score was 14.7. Overall, the percentage of patients with a mTICI score of 0, 2b, and 3 was 24.6, 26.2, and 49.2, respectively. Successful recanalization and first-pass success rates were 75.4% and 37.7%, respectively. There were no statistical differences between the results of the SUT and RID techniques. No vessel rupture, dissection, or device detachment was observed. The number of patients with a good clinical outcome (modified Rankin Scale score 0–2) was 17 (27.9%).ConclusionOur results showed that the Tigertriever device is safe and efficient for mechanical thrombectomy.
Databáze: OpenAIRE