Popis: |
A new external monitoring technique is described for measuring the amount of radioactivity within an internal organ. The variable distance between the detector and target organ or area is calculated trigonometrically by using two matched probes which are adjusted to register the same maximal level of activity coming from the source. The amount of radiation absorbed by the interposed tissue is determined from the ratio of transmitted counts of low energy nuclide such as I 125 , and a high energy nuclide as I 131 . Counting efficiency, i.e. the distance-radioactivity relationship, is constant for a specified detector-recording system and must be determined experimentally. The formula for calculating the microcuries of activity within an organ source is presented. Eine neue externe Strahlungspruftechnik zur Messung der Radioaktivitat innerhalb eines Organs wird beschrieben. Die variable Entfernung zwischen dem Strahlungsdetektor und dem Zielorgan oder-gebiet wird durch Verwendung zweier gleicher Meβeinrichtungen, die so eingestellt werden, daβ sie die gleiche maximale Strahlungsmenge von der Strahlungsquelle empfangen, trigonometrisch bestimmt. Die Menge der von dem dazwischen befindlichen Gewebe aufgenommenen Strahlung wird nach dem Verhaltnis der Zahlergebnisse der Strahlung eines Kerns niedriger Enrgie wie I 125 und ienes Kerns hoher Energie wie I 131 bestimmt. Die Zahlausbeute, d.h. das Verhaltnis Entfernung zu Radioaktivitat, ist konstant fur ein gegebenes Detektor-Aufzeichnungs-System und muβ experimentell bestimmt werden. Die Formel fur die Berechung der Mikrocuriezahl der Aktivitat innerhalb eines strahlenden Organs wird angegeben. |