Detection of bacteriuria by bioluminescence: effect of pre-analysis centrifugation of specimens

Autor: H.H. Johnston, Wright W. Nichols, G.D.W. Curtis
Rok vydání: 1984
Předmět:
Zdroj: Journal of Applied Bacteriology. 56:247-257
ISSN: 0021-8847
DOI: 10.1111/j.1365-2672.1984.tb01345.x
Popis: Nichols, W.W., Curtis, G.D.W. & Johnston, H.H. 1984. Detection of bacteriuria by bioluminescence: effect of pre-analysis centrifugation of specimens. Journal of Applied Bacteriology56, 247–257. Three bioluminescence-based, rapid methods of detecting significant bacteriuria were applied in parallel to 514 urine specimens. The results were compared with those of a quantitative pour plate viable count method, defined as positive if ≥ 105 c.f.u./ml of urine were observed. When adjusted to yield 21% falsely positive results the three rapid methods yielded 24%, 21% and 19% falsely negative results. If specimens with evidence of urethral or vaginal contamination were excluded (237 specimens remaining) the three methods yielded respectively 14%, 8% and 13% falsely negative results. A major source of disagreement between the bioluminescence-based methods and quantitative culture thus appeared to be contaminated urine specimens.