Determination of cyanide in waste waters using a quartz crystal microbalance
Autor: | João Pedro Oliveira, M. Teresa S.R. Gomes, Armando C. Duarte, A. Alexandre F. Silva |
---|---|
Předmět: |
Chemistry
Cyanide Metals and Alloys Analytical chemistry Quartz crystal microbalance Condensed Matter Physics Surfaces Coatings and Films Electronic Optical and Magnetic Materials Sample volume chemistry.chemical_compound Materials Chemistry Electrical and Electronic Engineering Electroplating Instrumentation |
Zdroj: | Scopus-Elsevier CIÊNCIAVITAE |
Popis: | A method based on a quartz crystal microbalance (QCM) was used for the first time to analyse cyanide in real samples. The method was applied to the analysis of waste waters from an electroplating industry. The analysis of replicates of five different samples ranging from 0.931 to 853 mg l −1 showed relative standard deviations between 1.7% and 9.3%. There was no evidence of systematic differences between the two sets of results obtained by the QCM method or by the ASTM method. However, the QCM method was faster and needed a sample volume 100 times smaller than the ASTM method. |
Databáze: | OpenAIRE |