P-Wave Sensing in VVI Pacemakers: Useful or a Problem?
Autor: | G.H.J. Tielen, M. I. H. El Gamal, L.M. Gelder |
---|---|
Rok vydání: | 1988 |
Předmět: |
Pacemaker
Artificial medicine.medical_specialty Ventricular lead medicine.diagnostic_test business.industry Heart Ventricles P wave General Medicine Vvi pacemaker Ventricular pacing Electrocardiography Heart Block Normal position Internal medicine medicine Cardiology Humans Female cardiovascular diseases Cardiology and Cardiovascular Medicine business Aged |
Zdroj: | Pacing and Clinical Electrophysiology. 11:1413-1418 |
ISSN: | 1540-8159 0147-8389 |
Popis: | Total inhibition of a bipolar VVI pacemaker was observed in an 84-year-old female after programming the pulse generator to its highest sensitivity (0.6 mV). After a short period of observation it became obvious that P-wave oversensing was responsible for this phenomenon. Frontal and lateral X-rays showed a normal position of the ventricular lead. Although oversensing could easily be avoided by decreasing sensitivity, changing the pacing mode to VVT converted the VVI pacemaker to an atrial synchronous ventricular pacing system. |