Characterisation of the PXIE Allison-type emittance scanner
Autor: | Lionel Prost, R. D׳Arcy, J. Gaynier, Alexander Shemyakin, M. Alvarez, V. Scarpine |
---|---|
Rok vydání: | 2016 |
Předmět: |
010302 applied physics
Physics Nuclear and High Energy Physics Scanner 010308 nuclear & particles physics business.industry Particle accelerator 01 natural sciences law.invention Pulse (physics) Optics law Error analysis Phase space 0103 physical sciences Calibration Thermal emittance Electronics business Instrumentation |
Zdroj: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 815:7-17 |
ISSN: | 0168-9002 |
Popis: | An Allison-type emittance scanner has been designed for PXIE at FNAL with the goal of providing fast and accurate phase space reconstruction. The device has been modified from previous LBNL/SNS designs to operate in both pulsed and DC modes with the addition of water-cooled front slits. Extensive calibration techniques and error analysis allowed confinement of uncertainty to the |