Is the atrial high rate episode diagnostic feature reliable in detecting paroxysmal episodes of atrial tachyarrhythmias?

Autor: Bernd Hauer, Jochen Senges, Eddy J. Vanagt, Karlheinz Seidl, Mike Hess, Ralf Zahn, Frans Ottenhoff, Eckard Meisel
Rok vydání: 1998
Předmět:
Zdroj: Pacing and clinical electrophysiology : PACE. 21(4 Pt 1)
ISSN: 0147-8389
Popis: The atrial high rate episode diagnostic in The Thera® pacemaker reports frequency, duration, and date/time of atrial tachyarrbytbmias according to programmed criteria. The aim of The study was to validate The atrial high rate episode diagnostic feature. Episodes of atrial fibrillation recorded by Holter monitoring were compared to episodes detected by the pacemaker. Eorty five ambulatory (Holter) recordings were used for evaluation. Thirty of 45 ambulatory (Holter) recordings showed sinus rhythm. On 4 of these 30 ambulatory (Holter) recordings, The Thera® detected 12 episodes of atrial tachyarrbythmias as false-positives (sinus rhythm was detected as atrial tacbyarrbytbmia). The main reason was far-field R and T wave oversensing. On 15 of 45 ambulatory (Holter) recordings, 125 episodes of atrial tachyarrhytbmias were recorded. Ninety-three of these events also were detected by the pacemaker, while for 32 events the Thera® reported sinus rhythm. The main reason was that the episodes were of too short duration. Therefore, the Thera® (programmed with detection rate 160 beats/min, detection beats 40, termination beats 10) was unable to detect atrial tachycardias. Software simulation of The diagnostic algorithm under several programming settings using the digitized Holter files demonstrated highly reliable detection of atrial tachyarrhythmias (sensitivity 98%, specificity 100%) when programmed as follows: detection rate 220 beats/min, detection beats 10, termination beats 20. It can be concluded that Thera®'s high rate episode monitor is a reliable tool for detection of a trial tachyarrhythmias, if programmed as recommended.
Databáze: OpenAIRE