Intrathecal Drug Delivery Systems (IDDS): The Implantable Systems Performance Registry (ISPR)
Autor: | Robert J. Plunkett, Eric Grigsby, Michelle D. Wells, Lisa M. Stearns, Todd Weaver, E. K. Stromberg, Mollie P. Roediger, John M. Huffman, Peter E. Konrad |
---|---|
Rok vydání: | 2016 |
Předmět: |
Adult
Male medicine.medical_specialty registry Intrathecal 03 medical and health sciences 0302 clinical medicine intrathecal drug delivery medicine Humans pain 030212 general & internal medicine Longitudinal Studies Registries Injections Spinal Aged Retrospective Studies Analgesics business.industry Intrathecal Drug Administration Systems spasticity Drug infusion General Medicine Infusion Pumps Implantable Middle Aged Survival Analysis United States Surgery Catheter Anesthesiology and Pain Medicine Treatment Outcome Neurology Muscle Spasticity Emergency medicine neuromodulation Female Neurology (clinical) business 030217 neurology & neurosurgery |
Zdroj: | Neuromodulation |
ISSN: | 1525-1403 |
Popis: | Objectives The ISPR was initially created to monitor the product performance of Medtronic implanted intrathecal drug infusion and spinal cord systems available in the United States. Materials and Methods Data were collected from 50 representative sites implanting and following patients with intrathecal drug delivery systems across the United States between August 7, 2003 and January 31, 2014. Device performance over time was estimated using life table survival methods. Results Of the 6093 patients enrolled in the ISPR, 3405 (55.9%) were female and 2675 (43.9%) were male, and 13 (0.2%) did not provide gender data. The average age at enrollment was 52.9 years (SD =17.6 years) and average follow-up time was 29.6 months. Currently, the estimates of device survival from pump-related events exceed 90% for all pump models across the applicable follow-up time points. The majority of product performance events were catheter-related. At 5 years of follow-up, all applicable catheter models, with the exception of revised not as designed or grafted not as designed catheters, had greater than 81% survival from catheter-related events. Conclusions The ISPR is designed to serve as an ongoing source of system and device-related information with a focus on “real-world” safety and product performance. ISPR data continue to be used to guide future product development efforts aimed at improving product reliability and quality. |
Databáze: | OpenAIRE |
Externí odkaz: |