The lifetime of a linac monitor unit ion chamber
Autor: | Dominic DiCostanzo, Ashley J. Cetnar |
---|---|
Rok vydání: | 2021 |
Předmět: |
Monitor unit
Treatment field Radiation business.industry Nuclear engineering ion chamber Linear particle accelerator Kapton Ion monitor unit chamber Ionization chamber Radiation Oncology Physics Humans Environmental science Radiology Nuclear Medicine and imaging Particle Accelerators Radiometry business Instrumentation Quality assurance Retrospective Studies |
Zdroj: | Journal of Applied Clinical Medical Physics |
ISSN: | 1526-9914 |
Popis: | This study is the first to report the clinical lifetime of Varian Kapton sealed ion chambers as a retrospective review. The data have been analyzed using ion chamber gain values, daily quality assurance results, monthly quality assurance results, and delivered treatment field data were analyzed to comprehensively review trends. The data show the average lifetimes of the ion chambers from our institution, so other physicists can prepare for replacement. Additionally, we share our experience in performing quality assurance tests to calibrate and validate the radiation beam after ion chamber replacement. |
Databáze: | OpenAIRE |
Externí odkaz: |