Safety and Efficacy of Continuous Flush Systems for Arterial and Pulmonary Artery Catheters
Autor: | Justin S. Clark, Reed M. Gardner, Edward L. Bond |
---|---|
Rok vydání: | 1977 |
Předmět: |
Pulmonary and Respiratory Medicine
business.industry Continuous infusion Arteries Static pressure Pulmonary Artery Drip chamber Arterial catheter medicine.disease Air embolism Catheterization Embolism Anesthesia medicine.artery Pulmonary artery otorhinolaryngologic diseases medicine Embolism Air Humans Flushing Surgery medicine.symptom Cardiology and Cardiovascular Medicine business |
Zdroj: | The Annals of Thoracic Surgery. 23:534-538 |
ISSN: | 0003-4975 |
DOI: | 10.1016/s0003-4975(10)63697-1 |
Popis: | Air embolism and static pressure errors have been attributed to continuous infusion systems. Experiments show that the primary source of air bubbles in such a system is the drip chamber. A drip chamber that minimizes this problem was tested and is recommended. Pressure errors due to the flush system are shown to be clinically insignificant. Fast flushing through a central arterial catheter is shown to be a minimal hazard. |