Autor: |
David O. Findley, Bruce Forell |
Rok vydání: |
1985 |
Předmět: |
|
Zdroj: |
International Journal of Radiation Oncology*Biology*Physics. 11:1727-1730 |
ISSN: |
0360-3016 |
DOI: |
10.1016/0360-3016(85)90227-5 |
Popis: |
Documented problems with loose and inert Ir-192 seeds in nylon ribbons prompted us to develop a quality control program. lieveral problem cases are illustrated. A sample of ribbons from each shipment is measured in a radioisotope dose calibrator. Results of the measurements over a 46-month period are presented. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|