Use of fast X-ray film for low-fluence biological electron microscopy at 100 kV and 1000 kV
Autor: | Michael Marko, Murray Vernon King, Donald F. Parsons |
---|---|
Rok vydání: | 1980 |
Předmět: |
Histology
Yield (engineering) Materials science Low fluence business.industry Analytical chemistry X-ray Electrons Electron Image Enhancement Radiation Dosage Fluence Pathology and Forensic Medicine law.invention Microscopy Electron law Optoelectronics Electron microscope business Beam (structure) Order of magnitude Mammography |
Zdroj: | Journal of Microscopy. 118:127-134 |
ISSN: | 0022-2720 |
Popis: | SUMMARY A user evaluation has been made by electron microscopists of an X-ray film for routine electron microscopy. The recent improvements in mammographic X-ray films, with the attempt to reduce the patient dose required to produce a high-resolution mammogram, have resulted in some useful films for medium- and high-voltage electron microscopy. They can yield essential cytological information with a reduction of the electron fluence (exposure) applied to the specimen of more than an order of magnitude compared with conventional electron-microscope films. Their use is indicated in situations where beam damage is severe. |