Test Strip Meconium Screening for Cystic Fibrosis

Autor: Jane A. Lacey, Thomas R. Connell, W. Theodore Bruns, Kenneth E. Whisler
Rok vydání: 1977
Předmět:
Zdroj: Archives of Pediatrics & Adolescent Medicine. 131:71
ISSN: 1072-4710
DOI: 10.1001/archpedi.1977.02120140073013
Popis: • The Boehringer-Mannheim Corporation (BMC) strip test is extremely reliable in indicating an albumin content above 20 mg/gm of dried meconium. All infants born during one year in 14 Milwaukee area hospitals were tested. Of 16,224 newborns, two were diagnosed correctly as suffering from cystic fibrosis and two were missed. False-positive tests were obtained in 0.9% of infants (prematurity, melena, gastroschisis, and intrauterine infection). The strip test is, at present, the best available but not the perfect screening method for cystic fibrosis. ( Am J Dis Child 131:71-73, 1977)
Databáze: OpenAIRE