Comparison of performance of self-expanding and balloon-expandable transcatheter aortic valves

Autor: Hoda Hatoum, Milad Samaee, Janarthanan Sathananthan, Stephanie Sellers, Maximilian Kuetting, Scott M. Lilly, Abdul R. Ihdayhid, Philipp Blanke, Jonathon Leipsic, Vinod H. Thourani, Lakshmi Prasad Dasi
Rok vydání: 2021
Předmět:
Zdroj: JTCVS open. 10
ISSN: 2666-2736
Popis: To evaluate the flow dynamics of self-expanding and balloon-expandable transcatheter aortic valves pertaining to turbulence and pressure recovery. Transcatheter aortic valves are characterized by different designs that have different valve performance and outcomes.Assessment of transcatheter aortic valves was performed using self-expanding devices (26-mm Evolut [Medtronic], 23-mm Allegra [New Valve Technologies], and small Acurate neo [Boston Scientific]) and a balloon-expandable device (23-mm Sapien 3 [Edwards Lifesciences]). Particle image velocimetry assessed the flow downstream. A Millar catheter was used for pressure recovery calculation. Velocity, Reynolds shear stresses, viscous shear stress, and pressure gradients were calculated.The maximal velocity at peak systole obtained with the Evolut R, Sapien 3, Acurate neo, and Allegra was 2.12 ± 0.19 m/sec, 2.41 ± 0.06 m/sec, 2.99 ± 0.10 m/sec, and 2.45 ± 0.08 m/sec, respectively (Flow dynamics downstream of different transcatheter aortic valves vary significantly depending on the valve type, despite not having a general trend depending on whether or not valves are self-expanding or balloon-expandable. Deployment design did not have an influence on flow dynamics.
Databáze: OpenAIRE