Feasibility and Usability of Patch-based Continuous Cardiac Rhythm Monitoring in Comparison with Traditional Telemetry in Noncritically Ill Hospitalized Patients
Autor: | John Tote, Alicia Burkle, Desiree Kirkwood, Ram Amuthan, Steven Mould, Molly Loy, Josalyn Meyer, Daniel J. Cantillon, Aaron C Hamilton, Shannon Pengel |
---|---|
Rok vydání: | 2019 |
Předmět: |
Alarm fatigue
medicine.medical_specialty business.industry cardiac telemetry Cardiac telemetry Vital signs Atrial fibrillation Usability medicine.disease Intensive care unit law.invention patch monitoring continuous cardiac monitoring law Physiology (medical) Telemetry Cardiac rhythm monitoring Ambulatory Emergency medicine Medicine Cardiology and Cardiovascular Medicine business Original Research |
Zdroj: | The Journal of Innovations in Cardiac Rhythm Management |
ISSN: | 2156-3993 2156-3977 |
DOI: | 10.19102/icrm.2019.100901 |
Popis: | Research on traditional cardiac telemetry demonstrates that excessive alarms are related to lead failures and noise-related interruptions. Patch-based continuous cardiac rhythm monitoring (CCRM) has emerged in outpatient ambulatory monitoring situations as a means to improve recording fidelity. In this study, patients hospitalized but not in the intensive care unit were simultaneously monitored via telemetry in parallel with the use of the Vital Signs Patch™ (VSP) CCRM system (LifeWatch Services, Rosemont, IL, USA), applying standardized monitoring and notifications provided by an off-site central monitoring unit (CMU). Among 11 patients (55% male; age: 66.8 ± 12.5 years), there were 42 CMU detections and 98 VSP detections. The VSP device was successfully applied by nursing with connectivity established in all 11 patients (100%). There were no VSP device–related adverse events or skin eruptions during the study. The CMU agreed with 59 (60%) of 98 VSP detections. Among those detections marked by disagreement 30 (77%) of 39 VSP detections were related to clinically meaningful arrhythmias (atrial: n = 9; ventricular: n = 7; brady-: n = 14) undetected by VSP due to noise. In two patients (18%), there were four clinically meaningful atrial fibrillation detections not recorded by the CMU. In conclusion, patch-based CCRM requires further development and review to replace traditional cardiac telemetry monitoring but could evolve into an appropriate method to detect clinically meaningful events missed by traditional methods if noise issues can be mitigated. |
Databáze: | OpenAIRE |
Externí odkaz: |