Drop-delay measurement using enzyme-coated particles

Autor: Andrey I. Poletaev, V. V. Zenin, Yuri Lazebnik
Rok vydání: 1992
Předmět:
Zdroj: Cytometry. 13:649-652
ISSN: 1097-0320
0196-4763
DOI: 10.1002/cyto.990130614
Popis: A simple technique is suggested for the measurement of drop delay for flow sorting. While the flow cytometer was set to sort a fixed number of particles, the drop-delay setting was changed step by step, and at each step the HRP-coupled particles were sorted into a well of an immunoassay strip. Then the HRP activity of the sorted samples was revealed by routine methods. The maximum level of the enzyme activity shows the proper drop-delay setting. Determination of the drop-delay setting takes only a few minutes. The technique is independent of the type of flow cytometer and does not require any additional equipment.