Two unusual cases of third degree skin burns are reported using MRI approved electrocardiographic leads. This is very uncommon as it is most often the electrodes which are the source of heat related issues. Both patients were sedated due to pain related issues of their lower spine. The burns were caused by a combination of using a 3 Tesla MRI scanner and the inability to cry out during scanning. We would like to bring forward a message that even when using MRI conditional equipment, clinical staff must be extremely careful in order to secure safe image acquisition using MRI. |