CLARION2-TRINITY: A Compton-suppressed HPGe and GAGG:Ce-Si-Si array for absolute cross-section measurements with heavy ions

Autor: T.J. Gray, J.M. Allmond, D.T. Dowling, M. Febbraro, T.T. King, S.D. Pain, D.W. Stracener, S. Ajayi, J. Aragon, L. Baby, P. Barber, C. Benetti, S. Bhattacharya, R. Boisseau, J. Gibbons, S.L. Tabor, V. Tripathi, C. Wibisono, I. Wiedenhöver, L. Bignell, M.S.M. Gerathy, G. Lane, L.J. McKie, A.J. Mitchell, J. Pope, R. du Rietz, A.E. Stuchbery, K. Vaigneur, T.J. Ruland
Rok vydání: 2022
Předmět:
Zdroj: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 1041:167392
ISSN: 0168-9002
Popis: The design and performance of a new Compton-suppressed HPGe and charged-particle array, CLARION2-TRINITY, are described. The TRINITY charged-particle array is comprised of 64 Cerium-doped Gadolinium Aluminium Gallium Garnet (GAGG:Ce) crystals configured into five rings spanning 7-54 degrees, and two annular silicon detectors that can shadow or extend the angular coverage to backward angles with minimal $\gamma$-ray attenuation. GAGG:Ce is a non-hygroscopic, bright, and relatively fast scintillator with a light distribution well matched to SiPMs. Count rates up to 40 kHz per crystal are sustainable. Fundamental characteristics of GAGG:Ce are measured and presented, including light- and heavy-ion particle identification (PID) capability, pulse-height defects, radiation hardness, and emission spectra. The CLARION2 array consists of up to 16 Compton-suppressed HPGe Clover detectors ($\approx4\%$ efficiency at 1 MeV) configured into four rings (eight HPGe crystal rings) using a non-Archimedean geometry that suppresses back-to-back coincident 511-keV gamma rays. The entire array is instrumented with 100- and 500-MHz (14 bit) waveform digitizers which enable triggerless operation, pulse-shape discrimination, fast timing, and pileup correction. Finally, two examples of experimental data taken during the commissioning of the CLARION2-TRINITY system are given: a PID spectrum from $^{16}$O + $^{18}$O fusion-evaporation, and PID and Doppler-corrected $\gamma$-ray spectra from $^{48}$Ti + $^{12}$C Coulomb excitation.
Comment: 14 pages, 21 figures
Databáze: OpenAIRE