Thickness of Molybdenum Filter and Squared Contrast-to-Noise Ratio per Dose for Digital Mammography
Autor: | Raleigh F. Johnson, Xizeng Wu, Thomas K. Nishino |
---|---|
Rok vydání: | 2005 |
Předmět: |
Digital mammography
Breast Neoplasms Radiation Dosage Imaging phantom Contrast-to-noise ratio Humans Medicine Dosimetry Mammography Radiology Nuclear Medicine and imaging Radiometry skin and connective tissue diseases Digital radiography Molybdenum medicine.diagnostic_test Phantoms Imaging business.industry Carcinoma Ductal Breast General Medicine Radiographic Image Enhancement Signal-to-noise ratio (imaging) Filter (video) Female Nuclear medicine business Monte Carlo Method |
Zdroj: | American Journal of Roentgenology. 185:960-963 |
ISSN: | 1546-3141 0361-803X |
DOI: | 10.2214/ajr.04.1489 |
Popis: | The objective of our study was to test whether the lesion-tissue contrast-to-noise ratio (CNR) at a given dose level can be improved by increasing the thickness of the molybdenum (Mo) filter currently used in digital mammography.We studied how the CNR between breast and a 5-mm simulated infiltrating ductal carcinoma (IDC) embedded in a 5-cm-thick breast changes with Mo filter thickness. We performed phantom imaging experiments by modifying the filter wheel of a Senographe 2000D unit with Mo filters that ranged from 15 to 90 microm in thickness. A 5-cm-thick 50% glandular-50% adipose breast phantom with a 5-mm insert simulating IDC was used as the phantom for all the cases. The CNRs between the breast phantom and the IDC insert were measured, and average glandular doses were calculated using a filtration-dependent X-ray spectra model and a breast dosimetry model based on a validated Monte Carlo simulation.The lesion-tissue CNR at a given dose level increases with increasing Mo filter thickness from 15 to 90 microm. The measured squared CNR per dose increased by 8%, 14%, 17%, and 17% for 45-, 60-, 75-, and 90-microm Mo filters, respectively, compared with the standard 30-microm Mo filter. Meanwhile, the exposure times were increased by 35% (45 microm), 71% (60 microm), 177% (75 microm), and 229% (90 microm).Increasing Mo filter thickness from 30 to 60 microm can increase lesion-tissue squared CNR per dose by 14% with a tolerable increase in the duration of exposure. |
Databáze: | OpenAIRE |
Externí odkaz: |