Development of a high-sensitivity radon monitor for a radon calibration system at KRISS
Autor: | Seung-Mi Hwang, Y.G. Seon, M.J. Han, J.M. Lee, K.B. Lee |
---|---|
Rok vydání: | 2020 |
Předmět: |
Radiation
Spectrometer Static Electricity Detector chemistry.chemical_element Radon 010403 inorganic & nuclear chemistry 01 natural sciences 030218 nuclear medicine & medical imaging 0104 chemical sciences 03 medical and health sciences 0302 clinical medicine chemistry Limit of Detection Calibration Environmental science Radiometry Sensitivity (electronics) Remote sensing |
Zdroj: | Applied Radiation and Isotopes. 164:109228 |
ISSN: | 0969-8043 |
DOI: | 10.1016/j.apradiso.2020.109228 |
Popis: | A radon calibration system is being developed with a high-sensitivity radon detector using electrostatic collection and an alpha spectrometer. Eight radon cells, each with a volume of 1.354 L, are integrated to achieve a high-sensitivity radon monitor with a total detector volume of 10.832 L. The sensitivity is estimated to be 35.6 ± 0.2 (36.8 ± 0.2) times higher relative to the RAD7 radon monitor for the detection of 218Po (214Po), respectively. |