Development of a high-sensitivity radon monitor for a radon calibration system at KRISS

Autor: Seung-Mi Hwang, Y.G. Seon, M.J. Han, J.M. Lee, K.B. Lee
Rok vydání: 2020
Předmět:
Zdroj: Applied Radiation and Isotopes. 164:109228
ISSN: 0969-8043
DOI: 10.1016/j.apradiso.2020.109228
Popis: A radon calibration system is being developed with a high-sensitivity radon detector using electrostatic collection and an alpha spectrometer. Eight radon cells, each with a volume of 1.354 L, are integrated to achieve a high-sensitivity radon monitor with a total detector volume of 10.832 L. The sensitivity is estimated to be 35.6 ± 0.2 (36.8 ± 0.2) times higher relative to the RAD7 radon monitor for the detection of 218Po (214Po), respectively.
Databáze: OpenAIRE