Standing Versus Supine Radiographs to Evaluate Femoral Head Penetration in the Polyethylene Liner After Total Hip Arthroplasty
Autor: | Keiji Tensho, Masaaki Maruyama, Shinji Wakabayashi, Kenji Hisa |
---|---|
Rok vydání: | 2014 |
Předmět: |
Adult
Male Ceramics medicine.medical_specialty Supine position Conventional polyethylene Arthroplasty Replacement Hip Radiography Posture Biocompatible Materials Young Adult Femoral head medicine Humans Orthopedics and Sports Medicine Aged Aged 80 and over business.industry Polyethylene liner Femur Head Penetration (firestop) Middle Aged Biocompatible material Prosthesis Failure Surgery medicine.anatomical_structure Polyethylene Female Hip Prosthesis Nuclear medicine business Total hip arthroplasty |
Zdroj: | The Journal of Arthroplasty. 29:2415-2419 |
ISSN: | 0883-5403 |
DOI: | 10.1016/j.arth.2014.02.010 |
Popis: | The linear penetration rates for ceramic femoral heads on conventional and highly cross-linked polyethylene were measured and compared with radiographs taken in the standing and supine position taken three weeks postoperatively and at final follow-up. Seventy-five patients (83 hips) with conventional polyethylene (group-1) and two hundred and seventy five patients (300 hips) with highly cross-linked polyethylene (group-2) were studied. Follow-up periods were 14.5years in group-1 and 8.6years in group-2. The average penetration rates in group-1 were 0.17mm/year in supine position and 0.18mm/year in standing position (P |
Databáze: | OpenAIRE |
Externí odkaz: |