Comparison of Sarns 3M heparin bonded to Duraflo II and control circuits in a porcine model: macro- and microanalysis of thrombi accumulation in circuit arterial filters

Autor: Leigh B. Kleinert, D A Arzouman, Douglas F. Larson, Vangie B. Patula, Stuart K. Williams
Rok vydání: 2000
Předmět:
Zdroj: Perfusion. 15:13-20
ISSN: 1477-111X
0267-6591
DOI: 10.1177/026765910001500103
Popis: Heparin-bonded perfusion circuits have been reported to reduce the thrombus formation during various levels of systemic heparinization. The goal of this study was to compare the efficacy of thrombo-resistance of the Sarns 3M heparin-bonded circuit to Baxter Duraflo II and untreated control in a porcine model. Fifteen Yorkshire pigs (60-65 kg) were anesthetized, heparinized with 3000 IU, intravenously (i.v.) and surgically cannulated with an internal jugular outflow and a femoral vein inflow. All circuits consisted of a 22-Fr venous cannula, centrifugal pump, arterial filter, an 18-Fr cannula for return and connected with equal lengths of 3/8″ polyvinyl chloride tubing. The flows were maintained at 2.0 l/min for 4 h. Thrombus formation in filter samples were morphometrically analyzed through macro-densitometry, light microscopy, and scanning electron microscopy (SEM). Our findings revealed that the 3M circuit had significantly less gross thrombus ( p < 0.001), 66% and 84% less microscopic thrombi and fivefold less SEM-measured aggregates ( p = 0.03) compared to the Duraflo II and uncoated groups. This study demonstrated that the 3M heparin-bonded circuit had significantly reduced the formation of micro- and macro-thrombi in the minimally heparinized pig model compared to the Duraflo II and untreated control circuits.
Databáze: OpenAIRE
Nepřihlášeným uživatelům se plný text nezobrazuje