Popis: |
1-Hydroxypyrene (1-OHP), a metabolite of polycyclic aromatic hydrocarbons (PAHs), is a frequently used biomarker for assessing human exposure to PAHs. Therefore, the technology that provides a quick, simple, cost-effective, portable, accurate, precise, and reliable test is still in great demand. To the best of our knowledge, the creation of an electrochemical device based on poly(l-glutamic acid)-modified a screen-printed graphene electrode (poly(L-GA)/SPGE) for 1-OHP detection was described for the first time. The developed sensor was simply and rapidly manufactured via only a single step of electropolymerization. All the concerned parameters and electroanalytical conditions were studied to obtain the best performance of the methodology. Under optimal conditions, the 1-OHP sensing provided a linear range of 1-1000 nM with the limits of detection and quantification of 0.95 and 3.16 nM, respectively. Moreover, this developed sensor was successfully utilized by determining 1-OHP in human urine samples. In comparison with conventional methods, this newly proposed electrochemical methodology might be tremendously valuable for 1-OHP evaluation in environmental and occupational applications, leading to the early detection of illness risk linked to PAHs in the human body. |