Flattening filter free beam energy selection and its impact in multitarget intracranial stereotactic radiosurgery treatments
Autor: | Parima Daroui, Hanako Farol, Nilam Ramsinghani, S Dietrich, Jeffrey V. Kuo, S. Laoui, D Roa, Jonathan Gonzales, Maxim Baldytchev |
---|---|
Rok vydání: | 2020 |
Předmět: |
Organs at Risk
Flattening filter free Radiological and Ultrasound Technology business.industry medicine.medical_treatment Radiotherapy Planning Computer-Assisted Planning target volume Radiotherapy Dosage Radiosurgery Volumetric modulated arc therapy Radiation therapy Multileaf collimator Oncology medicine Dosimetry Humans Radiology Nuclear Medicine and imaging Radiotherapy Intensity-Modulated Nuclear medicine business Beam energy |
Zdroj: | Medical dosimetry : official journal of the American Association of Medical Dosimetrists. 45(4) |
ISSN: | 1873-4022 |
Popis: | Extensive work has been done on the characteristics, dosimetry, and efficacy of flattening filter free (FFF) beams in radiosurgery. However, no study has addressed the dosimetric impact of FFF beam energy selection on treatment plan quality. This study aims to present a systematic dosimetric comparison of plan quality between 10 FFF vs 6 FFF beams in intracranial stereotactic radiosurgery (SRS) treatments using volumetric modulated arc therapy (VMAT). The dosimetric evaluation is based on radiation therapy oncology group (RTOG) dose conformity (CIRTOG) and gradient (GIRTOG) indices, and irradiated normal brain tissue volume. Thirty-five VMAT-based intracranial SRS treatments to multiple brain metastases using a 2.5 mm multileaf collimator (MLC) and 10 MV FFF beam were replanned with a 6 MV FFF and same MLC. The replans incorporated the same arc arrangement, planning target volume (PTV) and organs at risk structures, PTV coverage and prescription isodose normalization. The 6 MV FFF had a sharper dose fall-off compared to the 10 MV FFF (GIRTOG-10 FFF = 4.70, GIRTOG-6 FFF = 4.56; p |
Databáze: | OpenAIRE |