Validation of a method for estimating peak skin dose from CT-guided procedures

Autor: A. Kyle Jones, Meghan E. Kisiel, X. John Rong, Alda L. Tam
Rok vydání: 2021
Předmět:
Zdroj: Journal of Applied Clinical Medical Physics
ISSN: 1526-9914
Popis: A method for estimating peak skin dose (PSD) from CTDIvol has been published but not validated. The objective of this study was to validate this method during CT‐guided ablation procedures. Radiochromic film was calibrated and used to measure PSD. Sixty‐eight patients were enrolled in this study, and measured PSD were collected for 46 procedures. CTDIvol stratified by axial and helical scanning was used to calculate an estimate of PSD using the method [1.2 × CTDIvol(helical) + 0.6 × CTDIvol(axial)], and both calculated PSD and total CTDIvol were compared to measured PSD using paired t‐tests on the log‐transformed data and Bland‐Altman analysis. Calculated PSD were significantly different from measured PSD (P