Technical Note: A new TLD-phantom measurement system for determining dose distribution levels in the right and left breast from spiral CT chest imaging

Autor: Edgar Surprenant, Jeffery L. Hall, Jack Sklansky, Jorge L. Navarrete, Jack I. Eisenman
Rok vydání: 2002
Předmět:
Zdroj: Journal of Applied Clinical Medical Physics
ISSN: 1526-9914
DOI: 10.1120/jacmp.v3i4.2558
Popis: Two specially designed plastic/aluminum phantoms positioned thermoluminescent dosimeters (TLDs) at the right and left breast location of an anthrophomorophic chest torso. Imaging was performed on a spiral CT for a volume of the chest phantom through the breast area for a noncontiguous (pitch 1.5) helical chest scan. Conventional pencil beam ionization chamber measurements were made at the same operating parameters. The doses ranged from approximately 1 to 3 cGy. For both breast phantoms, the doses were highest for the medial inner quadrants near the mediastinum. The doses were lowest for the outer quadrants (lateral aspects) of both breasts. 2002 American College of Medical Physics.