A small-volume PVTX system for broadband spectroscopic calibration of downhole optical sensors

Autor: Jimmy W. Anders, Christopher Michael Jones, Jeffrey L. Moore, Jing Shen, Robert Atkinson, David L. Perkins, Michael T. Pelletier, Michael L. Myrick
Rok vydání: 2017
Předmět:
Zdroj: Review of Scientific Instruments. 88:073101
ISSN: 1089-7623
0034-6748
Popis: An instrument is presented that is capable of measuring the optical spectrum (long-wave ultraviolet through short-wave mid-infrared) of fluids under a range of temperature and pressure conditions from ambient pressure up to 138 MPa (20 000 psi) and 422 K (300 °F) using ∼5 ml of fluid. Temperature, pressure, and density are measured in situ in real-time, and composition is varied by adding volatile and nonvolatile components. The stability and accuracy of the conditions are reported for pure ethane, and the effects of temperature and pressure on characteristic regions of the optical spectrum of ethane are illustrated after correction for temperature and pressure effects on the optical cell path length, as well as normalization to the measured density. Molar absorption coefficients and integrated molar absorption coefficients for several vibrational combination bands are presented.
Databáze: OpenAIRE