Alumina reinforced zirconia implants: 1-year results from a prospective cohort investigation
Autor: | Benedikt Christopher Spies, Susanne Stampf, Markus Sperlich, Jonathan Fleiner, Ralf-Joachim Kohal |
---|---|
Rok vydání: | 2015 |
Předmět: |
medicine.medical_treatment
Radiography Alveolar Bone Loss Dentistry 02 engineering and technology Prosthesis Osseointegration 03 medical and health sciences Dental Implants Single-Tooth 0302 clinical medicine Aluminum Oxide medicine Humans Dental Restoration Failure Prospective Studies Prospective cohort study Gingival recession business.industry Soft tissue 030206 dentistry Single tooth 021001 nanoscience & nanotechnology Treatment Outcome Zirconium Implant Oral Surgery medicine.symptom 0210 nano-technology business Follow-Up Studies |
Zdroj: | Clinical Oral Implants Research. 27:481-490 |
ISSN: | 0905-7161 |
Popis: | Objective The objective of this investigation was to determine the clinical and radiographic outcome of a one-piece alumina-toughened zirconia (ATZ) oral implant for single tooth replacement after 1 year. Materials and Methods A total of 27 patients received one ATZ implant each in a one-stage implant surgery with immediate temporization. Standardized radiographs were taken at implant insertion and after 1 year to monitor peri-implant bone loss. Several soft tissue parameters were evaluated at prosthesis insertion and at the 1-year follow-up. Results Three patients did not receive a prosthetic reconstruction due to implant loss during the healing period. As a result, 24 patients were seen at the 1-year follow-up, giving a cumulative survival rate of 88.9%. The average marginal bone loss after 1 year was 0.77 mm. Only two implants (8.3%) lost at least 2 mm bone, whereas none of the implants lost more than 3 mm. Probing depth and clinical attachment level increased over 1 year at the implant sites, whereas gingival recession remained stable. Mombelli's bleeding and plaque indices showed no statistically significant change within the first year. Conclusion Owing to a total of 27 inserted implants, three failures within the first 4 months after implant placement carry weight. Therefore, the cumulative survival rate of the presented ceramic implant was slightly below reported survival rates of titanium implants when immediately restored. The recorded radiographic bone loss and peri-implant soft tissue parameters showed promising short-term results. The presented ATZ implant seems to be a candidate for clinical usage. |
Databáze: | OpenAIRE |
Externí odkaz: |