Negligible Specimen Hemolysis Is Observed With Sarstedt S-Monovette Coagulation Tubes in Aspiration Mode
Autor: | Abdulkadir Cat, Kamil Taha Ucar |
---|---|
Rok vydání: | 2022 |
Předmět: | |
Zdroj: | American Journal of Clinical Pathology. 158:494-498 |
ISSN: | 1943-7722 0002-9173 |
DOI: | 10.1093/ajcp/aqac071 |
Popis: | Objectives We aimed to compare the levels of hemolysis in the blood collected using the vacuum and aspiration modes via Sarstedt S-Monovette coagulation tubes. Methods Forty volunteers were included in the study. Blood samples were collected using two different modes in the S-Monovette citrate tube (Sarstedt AG). Prothrombin time, active partial thromboplastin time, fibrinogen, and D-dimer analyses were performed using the STA-Compact-Max 3 analyzer (Stago). The hemolysis levels of the samples were measured by both Stago’s semiquantitative hemolysis index (H-index) module and the quantitative H-index measurement of the Roche cobas 6000 (Roche Diagnostics) analyzer. Results Roche’s quantitative H-index values were statistically significantly lower in the aspiration mode. No clinically significant difference was observed between coagulation test results. Conclusions Using the S-Monovette citrate tubes can reduce spurious hemolysis and improve patient safety. |