Release of gas residues from an ethylene oxide sterilized haemodialyser

Autor: Vagn Handlos, Kurt Hoy
Rok vydání: 1983
Předmět:
Zdroj: Biomaterials. 4:321-322
ISSN: 0142-9612
DOI: 10.1016/0142-9612(83)90037-6
Popis: Ethylene oxide, a toxic gas used as sterilant for medical devices, is absorbed in plastic materials. The gas is desorbed from the sterilized devices occasionally leading to injuries to the users. A photoionization detector, being specific and sensitive, is used for measurement of liberated ethylene oxide from a haemodialyser. The detector, used as a flow through cell direct connected to the haemodialyser, enables measurement of ethylene oxide in quantities ten times smaller than the maximum daily exposure.
Databáze: OpenAIRE