Release of gas residues from an ethylene oxide sterilized haemodialyser
Autor: | Vagn Handlos, Kurt Hoy |
---|---|
Rok vydání: | 1983 |
Předmět: |
Ethylene Oxide
Materials science Ethylene oxide Radiochemistry Biophysics Plastic materials Flow through cell Analytical chemistry Sterilization Bioengineering Sterilization (microbiology) Toxic gas Biomaterials Photoionization detector chemistry.chemical_compound chemistry Mechanics of Materials Ceramics and Composites Daily exposure Radiometry Kidneys Artificial |
Zdroj: | Biomaterials. 4:321-322 |
ISSN: | 0142-9612 |
DOI: | 10.1016/0142-9612(83)90037-6 |
Popis: | Ethylene oxide, a toxic gas used as sterilant for medical devices, is absorbed in plastic materials. The gas is desorbed from the sterilized devices occasionally leading to injuries to the users. A photoionization detector, being specific and sensitive, is used for measurement of liberated ethylene oxide from a haemodialyser. The detector, used as a flow through cell direct connected to the haemodialyser, enables measurement of ethylene oxide in quantities ten times smaller than the maximum daily exposure. |
Databáze: | OpenAIRE |
Externí odkaz: |