Comparison of NIOSH 5040 Method versus Aethalometer™ to Monitor Diesel Particulate in School Buses and at Work Sites

Autor: G. Sirianni, Howard J. Cohen, S. Chemerynski, R. Wheeler, Jonathan Borak
Rok vydání: 2003
Předmět:
Zdroj: AIHA Journal. 64:260-268
ISSN: 1542-8117
DOI: 10.1080/15428110308984816
Popis: This study, undertaken initially to understand apparent differences between two recent reports of diesel particulate matter (DPM) inside school buses, involved side-by-side comparisons of two different methods for measuring ambient and occupational exposures to DPM. The NIOSH 5040 method yields time-weighted-average measurements of elemental carbon (EC) and organic carbon (OC), whereas the Aethalometer yields near-real time measures of carbonaceous particles (black carbon or BC) by optical absorption. The two methods were compared on three school buses and in three different work sites. Three side-by-side sampling configurations were used (open-faced filter with and without two different cyclones) in triplicate along with two or three Aethalometers. BC readings were correlated with results from open-faced cassettes, but there were statistically significant differences between the results of side-by-side Aethalometers. In addition, Aethalometer airflow rates were inaccurate, the instruments were sensitive to vibration, optical calibration could not be performed, and historically derived conversion factors to relate EC and BC were not appropriate. The Aethalometer could prove useful for monitoring of industrial work sites, but only after a method for external calibration and improvements in pump design and vibration isolation are developed.