Popis: |
Tramadol has extracted from the exhaled breath condensate (EBC) samples through the supported liquid membrane consisting of 2-nitrophenyl octyl ether impregnated in the hollow fiber wall, and the lumen of the hollow fiber was filled with 20 μL of an acceptor phase. Under the optimum conditions of the electromembrane extraction, i.e. the stirring speed of 750 rpm, extraction time of 20 min, acceptor pH at 1.0, donor phase pH at 6.0, and an applied voltage of 170 V across the supported liquid membrane, a preconcentration factor of 128-fold with a extraction recovery of 64% was achieved. Acceptable linearity was obtained in the tramadol concentration range of 5-1000 ng mL |