Pacemaker electrode repositioning using the loop-snare technique
Autor: | Ian R. Scott, W.R. Eric Jamieson, D. Christopher Morris |
---|---|
Rok vydání: | 1989 |
Předmět: |
medicine.medical_specialty
Cardiac Catheterization Pacemaker Artificial Heart Ventricles Femoral vein Vena Cava Inferior Inferior vena cava medicine Humans cardiovascular diseases Lead (electronics) Aged business.industry General Medicine Surgery Ventricular pacemaker Electrodes Implanted Loop (topology) Catheter medicine.vein Fluoroscopy Electrode cardiovascular system Female Cardiology and Cardiovascular Medicine business Closed loop Biomedical engineering |
Zdroj: | Pacing and clinical electrophysiology : PACE. 12(6) |
ISSN: | 0147-8389 |
Popis: | Using a percutaneous femoral vein approach under fluoroscopic control, a malpositioned ventricular pacemaker electrode was released from the right ventricular vvall by hooking the lead with a deflecting wire inserted into a RIM catheter. A closed loop was formed by tightening the handle of the wire allowing the electrode to be dislodged and pulled into the inferior vena cava. The electrode was then snared using a loop formed hy an exchange wire advanced through an 8 French catheter with a J-curve steamed at its tip. The electrode was advanced to the right ventricular apex and released by advancing one end of the snare wire while pulling the other end to open the loop. |