Determination of human serum semicarbazide-sensitive amine oxidase activity via flow injection analysis with fluorescence detection after online derivatization of the enzymatically produced benzaldehyde with 1,2-diaminoanthraquinone

Autor: Takahiro Imazato, Mahmoud El-Maghrabey, Yukitaka Ueki, Kaname Ohyama, Naotaka Kuroda, Naoya Kishikawa
Rok vydání: 2015
Předmět:
Zdroj: Analytica Chimica Acta. 881:139-147
ISSN: 0003-2670
DOI: 10.1016/j.aca.2015.04.006
Popis: A fast, simple, and sensitive flow injection analysis method was developed for the measurement of semicarbazide-sensitive amine oxidase (SSAO) activity in human serum. Benzaldehyde, generated by the action of SSAO after incubation of serum with benzylamine, was derivatized with a novel aromatic aldehyde-specific reagent (1,2-diaminoanthraquinone) and the fluorescent product was measured by fluorescence detection at excitation and emission wavelengths of 390 and 570nm, respectively. Serum SSAO activity was defined as benzaldehyde (nmol) formed per milliliter serum per hour. The method was linear over SSAO activity of 0.2-150.0nmolmL-1h-1 with a detection limit of 0.06nmolmL-1h-1. The %RSD of intra-day and inter-day precision did not exceed 9.4% and the accuracy ranged from -6.5 to -0.6%. The method was applied for the determination of the serum SSAO activity in healthy controls (C, n=24) and diabetes mellitus patients (DM, n=18). It was demonstrated that the activity (mean±SE) of SSAO in diabetics sera was significantly higher than that in healthy subjects' ones (DM; 73.3±1.8nmolmL-1h-1 vs C; 58.9±2.2nmolmL-1h-1, P
Analytica Chimica Acta, 881, pp.139-147; 2015
Databáze: OpenAIRE