Dimensional Variations of Left-Sided Double-Lumen Endobronchial Tubes
Autor: | Alexander Dullenkopf, Sebastian Schnitzler, Niels Hegland, Markus Weiss, Marc P. Steurer, Jan Ellensohn |
---|---|
Přispěvatelé: | University of Zurich, Dullenkopf, Alexander |
Rok vydání: | 2019 |
Předmět: |
Article Subject
business.industry Endobronchial tubes 610 Medicine & health 030204 cardiovascular system & hematology Critical Care and Intensive Care Medicine Left sided lcsh:RD78.3-87.3 03 medical and health sciences 0302 clinical medicine Anesthesiology and Pain Medicine lcsh:Anesthesiology 030202 anesthesiology Clinical Research Cuff Medicine 10220 Clinic for Surgery 2703 Anesthesiology and Pain Medicine 2706 Critical Care and Intensive Care Medicine business Research Article Biomedical engineering Lumen (unit) |
Zdroj: | Anesthesiology Research and Practice, Vol 2019 (2019) Anesthesiology Research and Practice |
Popis: | Background. Tube size selection is critical in ventilating patients’ lungs using double-lumen endobronchial tubes (DLTs). Little information about relevant parameters is readily available from manufacturers. The aim of this study is to provide reference data for relevant dimensions of conventionally available DLTs. Methods. In this study in a benchmark in vitro setup, several dimensional parameters of four sizes of left-sided double-lumen endobronchial tubes from six different manufacturers were assessed, such as distances and diameters of tube shaft, cuff lengths, and diameters as well the angle at the tip. Results. Endobronchial tubes of ostensibly the same size revealed wide variation in measured parameters between brands from different manufacturers. In some parameters, there was an overlap between different sizes from the same manufacturer, i.e., diameters and distances did not increase with increasing nominal endobronchial tube size. The information about dimensions of endobronchial tubes provided by manufacturers’ leaflets is insufficient. Conclusions. Endobronchial tube size selection carries unnecessary uncertainty because clinically relevant parameters are unknown and vary considerably between different manufacturers. |
Databáze: | OpenAIRE |
Externí odkaz: |