Assessment of setup uncertainty in hypofractionated liver radiation therapy with a breath-hold technique using automatic image registration–based image guidance
Autor: | Sunil Krishnan, Joseph M. Herman, Eugene J. Koay, Gye Won Choi, Bruce D. Minsky, Sam Beddar, Albert C. Koong, Grace L. Smith, Prajnan Das, Cullen M. Taniguchi, Yelin Suh, Emma B. Holliday |
---|---|
Rok vydání: | 2019 |
Předmět: |
lcsh:Medical physics. Medical radiology. Nuclear medicine
Setup uncertainty Hypofractionated Radiation Therapy lcsh:R895-920 medicine.medical_treatment Residual lcsh:RC254-282 Liver radiotherapy Breath Holding Automatic image registration 03 medical and health sciences 0302 clinical medicine Image Processing Computer-Assisted medicine Humans PTV margin Radiology Nuclear Medicine and imaging IGRT Image-guided radiation therapy business.industry Radiotherapy Planning Computer-Assisted Research Liver Neoplasms Uncertainty Ptv margin Cone-Beam Computed Tomography lcsh:Neoplasms. Tumors. Oncology. Including cancer and carcinogens Prognosis In-room CT Radiation therapy Oncology 030220 oncology & carcinogenesis Breathing Measurement uncertainty Radiation Dose Hypofractionation Radiotherapy Intensity-Modulated Nuclear medicine business Radiotherapy Image-Guided |
Zdroj: | Radiation Oncology (London, England) Radiation Oncology, Vol 14, Iss 1, Pp 1-9 (2019) |
ISSN: | 1748-717X |
DOI: | 10.1186/s13014-019-1361-6 |
Popis: | Background Target localization in radiation therapy is affected by numerous sources of uncertainty. Despite measures to minimize the breathing motion, the treatment of hypofractionated liver radiation therapy is further challenged by residual uncertainty coming from involuntary organ motion and daily changes in the shape and location of abdominal organs. To address the residual uncertainty, clinics implement image-guided radiation therapy at varying levels of soft-tissue contrast. This study utilized the treatment records from the patients that have received hypofractionated liver radiation therapy using in-room computed tomography (CT) imaging to assess the setup uncertainty and to estimate the appropriate planning treatment volume (PTV) margins in the absence of in-room CT imaging. Methods We collected 917 pre-treatment daily in-room CT images from 69 patients who received hypofractionated radiation therapy to the liver with the inspiration breath-hold technique. For each treatment, the daily CT was initially aligned to the planning CT based on the shape of the liver automatically using a CT-CT alignment software. After the initial alignment, manual shift corrections were determined by visual inspection of the two images, and the corrections were applied to shift the patient to the physician-approved treatment position. Considering the final alignment as the gold-standard setup, systematic and random uncertainties in the automatic alignment were quantified, and the uncertainties were used to calculate the PTV margins. Results The median discrepancy between the final and automatic alignment was 1.1 mm (0–24.3 mm), and 38% of treated fractions required manual corrections of ≥3 mm. The systematic uncertainty was 1.5 mm in the anterior-posterior (AP) direction, 1.1 mm in the left-right (LR) direction, and 2.4 mm in the superior-inferior (SI) direction. The random uncertainty was 2.2 mm in the AP, 1.9 mm in the LR, and 2.2 mm in the SI direction. The PTV margins recommended to be used in the absence of in-room CT imaging were 5.3 mm in the AP, 3.5 mm in the LR, and 5.1 mm in the SI direction. Conclusions Manual shift correction based on soft-tissue alignment is substantial in the treatment of the abdominal region. In-room CT can reduce PTV margin by up to 5 mm, which may be especially beneficial for dose escalation and normal tissue sparing in hypofractionated liver radiation therapy. |
Databáze: | OpenAIRE |
Externí odkaz: |