A simple device to inject indicator gas for wash-out tests during mechanical ventilation

Autor: Wim P.J. Holland, P. E. M. Huygen, C. Jabaaij, Hajo A. Bruining, Can Ince, I. Gultuna
Rok vydání: 1992
Předmět:
Zdroj: Intensive Care Medicine. 18:304-308
ISSN: 1432-1238
0342-4642
Popis: Objective: To evaluate a simple device which injects a constant fraction of indicator gas to the inspiratory mixture for performing multi-breath wash-out tests during controlled ventilation.Design: the technique in which the indicator gas is injected at the mouth of the patient (post-mix) is compared with the technique where the indicator gas is administered in the bellows of the ventilator (pre-mix).Setting: Surgical Intensive Care Unit of a University Hospital.Patients: 10 post-operative mechanically ventilated patients.Interventions: None.Measurements and results: 3 wash-out tests with the post-mix and 3 wash-out tests with the pre-mix method were performed within an hour on every patient. The calculated mean end expiratory lung volume (EEV) was 1.91±0.871 with the post-mix technique and 1.89±0.881 with the premix technique. There was a good agreement with a mean difference of −1.9±6.5% in the calculated EEV values by the two different techniques.Conclusion: The described injector is an affordable device, is easy to assemble and can be incorporated in most electronically regulated ventilators to perform multi-breath indicator gas wash-out tests for pulmonary monitoring at the bed side of ICU patients.
Databáze: OpenAIRE