Effects of geometric and screen-film unsharpness in conventional and 350-kVp chest radiography
Autor: | J Meyer, A G Haus, L B North |
---|---|
Rok vydání: | 1980 |
Předmět: | |
Zdroj: | Radiology. 137:197-202 |
ISSN: | 1527-1315 0033-8419 |
DOI: | 10.1148/radiology.137.1.7422845 |
Popis: | Clinical comparison of 350- and conventional 120-kVp techniques indicates that small linear structures, such as small pulmonary blood vessels, are shown better at 120 kVp, due in part to two factors affecting image unsharpness: geometric and screen-film unsharpness. The combined effects of geometric (Ug) and screen-film unsharpness were evaluated for both kVp levels. The total-system modulation transfer function (MTFUg X MTFscreen-film) demonstrates differences in unsharpness under two conditions: (a) when the same screen-film system is used with different magnifications (i.e., extreme planes in the chest) and (b) when different screen-film systems are used for magnification of the midplane of the chest. |