Efficiency of Drug Delivery to the Coronary Arteries in Swine Is Dependent on the Route of Administration: Assessment of Luminal, Intimal, and Adventitial Coronary Artery and Venous Delivery Methods
Autor: | Chul Ahn, Jennifer Peregoy, Tracy L. Murray, William F. Pritchard, John W. Karanian, O. Alberto Chiesa |
---|---|
Rok vydání: | 2010 |
Předmět: |
medicine.medical_specialty
Lung Swine business.industry Anterior Descending Coronary Artery Balloon Coronary Vessels Coronary arteries Route of administration Catheter Drug Delivery Systems medicine.anatomical_structure Albumins Internal medicine Drug delivery medicine Cardiology Animals Infusions Intra-Arterial Radiology Nuclear Medicine and imaging Infusions Intravenous Cardiology and Cardiovascular Medicine business Artery |
Zdroj: | Journal of Vascular and Interventional Radiology. 21:1555-1564 |
ISSN: | 1051-0443 |
Popis: | Purpose To compare the efficiency of five different drug delivery methods to the coronary artery in swine. Materials and Methods A nanoparticle-albumin-bound, nonradioactive isotopic marker was administered within the left anterior descending coronary artery (LAD) through a microinfusion catheter (MIC: adventitial, n = 8, and luminal, n = 4), a porous drug infusion balloon (DIB: intimal, n = 4), and a straight catheter (SC: luminal, n = 2) and within the superior vena cava (SC: intravenous, luminal, n = 2). The distribution of the marker in heart, lung, liver, kidney, muscle, blood, urine, and bile was determined 68–84 minutes after delivery. The heart was sectioned into six axial slices and each slice divided into four quadrants. The marker content was assayed by neutron bombardment and the total counts of disintegrations per minute (DPM) expressed as a percentage of the control for each device delivery control. Results After luminal delivery with the nonactuated MIC (MIC-NA) or intimal delivery with the DIB, 0.17% ± 0.07 and 0.39% ± 0.09, respectively, less than 0.39% of the total marker was detected in the heart. After adventitial delivery with the actuated MIC (MIC-A), 63.1% ± 9.9 of the total marker was detected in the heart. Marker was only detected in quadrants containing the coronary LAD, with the highest level in the middle slice and lower marker levels in consecutive proximal and distal heart slices. The nonactuated MIC-NA and DIB drug infusion balloon patterns of marker distribution were similar to those of actuated MIC-A, although with reduced levels. These delivery methods were also associated with considerably more marker detected in the lungs and liver: at least 22% compared with 1.34% ± 1.34 for the actuated MIC-A There was one delivery failure with the actuated MIC. Conclusions Catheter-based adventitial delivery with the MIC-A represents a more efficient delivery method for retention of vascular therapeutics. |
Databáze: | OpenAIRE |
Externí odkaz: |