Decontamination of filtering facepiece respirators in primary care using medical autoclave

Autor: Bernadette van Maltha-van Santvoort, Jonathan Bouman, Tim Horeman, Ralf E. Harskamp, Bart van Straten, John J. van den Dobbelsteen, Joost R.M. van der Sijp
Jazyk: angličtina
Rok vydání: 2020
Předmět:
DOI: 10.1101/2020.04.27.20066654
Popis: ObjectiveThere are widespread shortages of personal protective equipment as a result of the coronavirus disease 2019 (COVID-19) pandemic. Reprocessing filtering facepiece respirators may provide an alternative solution in keeping health care professionals safe.Designprospective, bench-to-bedsideSettingA primary care-based study using filtering facepiece particles (FFP) type 2 respirators without exhalation valve (3M Aura 1862+, Maco Pharma ZZM002), FFP2 respirators with valve (3M Aura 9322+ and San Huei 2920V), and valved FFP type 3 respirators (Safe Worker 1016).InterventionsAll masks were reprocessed using a medical autoclave (34-minute total cycle time of steam sterilization, with 17 minutes at 121°C) and subsequently tested up to 3 times whether these decontaminated respirators retained their integrity (seal check, pressure drop) and ability to filter small particles (0.3-5.0μm) in the laboratory using a particle penetration test.ResultsWe tested 32 respirators, and 63 samples for filter capacity. All 27 FFP-2 respirators retained their shape, whereas half of the sterilized FFP-3 respirators (Safe Worker 1116) showed deformities and failed the seal check. The filtering capacity of the 3M Aura 1862 was best retained after 1, 2, and 3 sterilization cycles (0.3μm: 99.3±0.3% (new) versus 97.0±1.3, 94.2±1.3% or 94.4±1.6, pConclusionThis study shows that selected FFP2-type respirators may be reprocessed for use in primary care, as the tested masks retain their shape, ability to retain particles and breathing comfort after decontamination using a medical autoclave.Strengths and limitations of this study-Pragmatic use of autoclave to sterilize and reuse filter facepiece respirators-Combining clinical and laboratory findings to evaluate the safety in terms of shape, ability to retain particles and breathing comfort-The study is limited in sample size and restricted to selected FFP-2 and FFP-3 respirators-The study is a first of its kind in primary care settings and thus unvalidated-The study does not provide "hard" clinical evidence in terms of a randomized trial (i.e. reprocessed mask versus usual care)
Databáze: OpenAIRE