Evaluation of a noise reduction imaging technology in iliac digital subtraction angiography: noninferior clinical image quality with lower patient and scatter dose

Autor: Maria Mauti, T. Grünhagen, Marc van Leersum, Peter Gaines, Jan B.A. Habraken, Graham J. Robinson, Hans van Overhagen, Marco J. van Strijen, Nicholas Railton, Markus Zähringer
Rok vydání: 2014
Předmět:
Zdroj: Journal of vascular and interventional radiology : JVIR. 26(5)
ISSN: 1535-7732
Popis: Purpose To determine whether equivalent-quality images can be obtained from digital subtraction angiography (DSA) of the iliac artery after implementation of a novel imaging technology that reduces patient and scatter x-ray dose. Materials and Methods Imaging using two randomly ordered DSA runs was performed in 51 adults scheduled for iliac artery angiography or intervention or both. One DSA run used standard acquisition chain and image processing algorithms (referred to as " reference DSA"), and the other DSA run used dose-reduction and real-time advanced image noise reduction technology (referred to as "study DSA"). The quality of each pair of runs, consecutively performed without changes in working projection or injection parameters, was independently rated by five radiologists blinded to the imaging technology used. Patient radiation dose was evaluated using air kerma and dose area product, and scatter dose was evaluated using three dosimeters (DoseAware, Philips Healthcare, Best, The Netherlands), located at fixed positions. Results Comparable image pairs were available in 48 patients. There were 44 patients undergoing treatment involving the common (n = 33) or external (n = 29) iliac arteries. Study DSA images were rated as equal to or better than reference DSA images for 96% of comparisons, with an average overall agreement among raters of 0.93 (95% confidence interval, 0.65–0.96). Mean patient radiation dose (n = 48) and scatter dose rate for the three dosimeters (n = 50) was 83% ± 5 and 69% ± 10 lower, respectively, using the study technology ( P Conclusions Iliac artery DSA performed using a dose-reduction and real-time advanced image noise reduction technology results in image quality that is noninferior to conventional DSA but with significantly lower patient and scatter radiation exposure ( P
Databáze: OpenAIRE