Popis: |
Magnetically controlled growing rods (MCGR) have become the dominant distraction-based implant for the treatment of early onset scoliosis (EOS). Recent studies, however, have demonstrated rising rates of implant failure beyond short-term follow-up. We sought to evaluate a single-center experience with MCGR for the treatment of EOS to define the rate of MCGR failure to lengthen, termed implant stall, over time.A single-center, retrospective review was conducted identifying children with EOS undergoing primary MCGR implantation. The primary endpoint was the occurrence of implant stalling, defined as a failure of the MCGR to lengthen on three consecutive attempted lengthening sessions with minimum of 2 years follow-up. Clinical and radiographic variables were collected and compared between lengthening and stalled MCGRs. A Kaplan-Meier survival analysis was conducted to assess implant stalling over time.A total of 48 children met inclusion criteria (mean age 6.3 ± 1.8 years, 64.6% female). After a mean 56.9 months (range of 27 to 90 months) follow-up, 25 (48%) of children experienced implant stalling at a mean of 26.0 ± 14.1 months post-implantation. Kaplan-Meier survival analysis demonstrated that only 50% of MCGR continue to successfully lengthen at 2 years post-implantation, decreasing to 20% at 4 years post-implantation.Only 50% of MCGR continue to successfully lengthen 2 years post-implantation, dropping dramatically to 20% at 4 years, adding to the available knowledge regarding the long-term viability and cost-effectiveness of MCGR in the management of EOS. Further research is needed to validate these findings. |