A feasibility study of UHPLC-HRMS accurate-mass screening methods for multiclass testing of organic contaminants in food
Autor: | Patricia Pérez-Ortega, Felipe J. Lara-Ortega, Marek Trojanowicz, Antonio Molina-Díaz, Bienvenida Gilbert-López, Juan F. García-Reyes |
---|---|
Přispěvatelé: | Universidad de Jaén, Ministerio de Economía y Competitividad (España) |
Rok vydání: | 2016 |
Předmět: |
Nitrosamines
Collision-induced dissociation Electrospray ionization Automated data processing Analytical chemistry Food Contamination Mass spectrometry 01 natural sciences Mass Spectrometry Analytical Chemistry Matrix (chemical analysis) Liquid chromatography–mass spectrometry Screening methods Pesticides Mass screening Chromatography High Pressure Liquid Fluorocarbons Chromatography 010405 organic chemistry Chemistry 010401 analytical chemistry Veterinary Drugs Mycotoxins 0104 chemical sciences LC-MS Mass Pesticide Food Sweetening Agents Feasibility Studies Infant Food |
Zdroj: | Digital.CSIC. Repositorio Institucional del CSIC instname |
ISSN: | 1873-3573 |
Popis: | The feasibility of accurate-mass multi-residue screening methods using liquid chromatography high-resolution mass spectrometry (UHPLC-HRMS) using time-of-flight mass spectrometry has been evaluated, including over 625 multiclass food contaminants as case study. Aspects such as the selectivity and confirmation capability provided by HRMS with different acquisition modes (full-scan or full-scan combined with collision induced dissociation (CID) with no precursor ion isolation), and chromatographic separation along with main limitations such as sensitivity or automated data processing have been examined. Compound identification was accomplished with retention time matching and accurate mass measurements of the targeted ions for each analyte (mainly (de)protonated molecules). Compounds with the same nominal mass (isobaric species) were very frequent due to the large number of compounds included. Although 76% of database compounds were involved in isobaric groups, they were resolved in most cases (99% of these isobaric species were distinguished by retention time, resolving power, isotopic profile or fragment ions). Only three pairs could not be resolved with these tools. In-source CID fragmentation was evaluated in depth, although the results obtained in terms of information provided were not as thorough as those obtained using fragmentation experiments without precursor ion isolation (all ion mode). The latter acquisition mode was found to be the best suited for this type of large-scale screening method instead of classic product ion scan, as provided excellent fragmentation information for confirmatory purposes for an unlimited number of compounds. Leaving aside the sample treatment limitations, the main weaknesses noticed are basically the relatively low sensitivity for compounds which does not map well against electrospray ionization and also quantitation issues such as those produced by signal suppression due to either matrix effects from coeluting matrix or from coeluting analytes present in the standards solutions which often occur as they contain hundreds of the analytes included in the screening methods. The authors acknowledge funding from the Spanish Ministerio de Economía y Competitividad (MINECO) (Refs. CTQ-2009-10897 and CTQ-2012-34297). P.P.-O acknowledges a PhD scholarship from University of Jaén. F.J.L.-O. acknowledges a PhD student scholarship from MINECO (Ref. BES-2013-064014). |
Databáze: | OpenAIRE |
Externí odkaz: |