Electromagnetic field Materials science Hot Temperature Resonant cavity Imaging phantom Body Temperature Heating Optics medicine Humans Computer Simulation Electrodes Human head business.industry Brain Neoplasms Phantoms Imaging Temperature Hyperthermia Treatment Equipment Design Hyperthermia Induced medicine.disease Agar Electrode Development (differential geometry) business Head Biomedical engineering |
Zdroj: | EMBC |
ISSN: | 1557-170X |
Popis: | A re-entrant type resonant cavity applicator for brain tumor hyperthermia treatment is presented. In this method, a human head is placed between the gap of the inner re-entrant cylinders, and is heated with electromagnetic fields stimulated in the cavity without contact between the surface of the human head and the applicator. First, the comparison of the computer simulation and experimental results were presented. The estimated temperature agrees with the measured temperature with an error of 10% or better. Second, a method to control the heating pattern was presented. In the method, a human head was moved towards the electrode. The controllable range of heating pattern was about 70% of the distance from the center of the agar phantom in the direction of depth. |