Agreement between Compass Fundus Perimeter New Grid and 10-2 Testing Protocols for Detecting Central Visual Field Defects
Autor: | Nevin W. El-Nimri, Sasan Moghimi, Robert N. Weinreb, Huiyuan Hou, Linda M. Zangwill, Elham Ghahari, Christopher Bowd, James A. Proudfoot, Rafaella C. Penteado, Patricia Isabel C. Manalastas |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2021 |
Předmět: |
medicine.medical_specialty
genetic structures business.industry Glaucoma General Medicine Fundus (eye) medicine.disease Confidence interval Article Visual field Perimeter Pattern standard deviation Cross-Sectional Studies Ophthalmology Compass medicine Humans Visual Fields business Scotoma Glaucoma Open-Angle Decibel |
Zdroj: | Ophthalmol Glaucoma |
Popis: | PURPOSE: To evaluate the agreement between Compass New Grid (NG) and 10-2 test protocols for detecting early glaucomatous defects in the central 10 degrees of the visual field (CVFD). DESIGN: Cross-sectional study. SUBJECTS AND CONTROLS: A total of 123 eyes of 14 healthy individuals, 17 glaucoma suspects and 32 glaucoma patients were enrolled. METHODS: Subjects performed NG and 10-2 Compass automated perimetry testing within one week. For both test protocols total deviation (TD) or pattern deviation (PD) plot CVFDs were defined by three contiguous points with probabilities of |
Databáze: | OpenAIRE |