Criteria of slope and endogenous level for selection of matrix in the quantitation of endogenous compounds by LC-MS/MS
Autor: | Allena Ji Ji, Guangxin Xu, Lin Tao, Hongqin Jenny Jiao, Rachel Elisabeth Palmer |
---|---|
Rok vydání: | 2022 |
Předmět: | |
Zdroj: | Bioanalysis. 14:807-816 |
ISSN: | 1757-6199 1757-6180 |
Popis: | Selection of the appropriate matrix for standard curve is critical for an accurate and sensitive biomarker method. Slope of a standard curve is the key factor for parallelism assessment between tested matrix and authentic matrix for LC-MS/MS assays. Here the authors have established slope criteria using a generic equation and endogenous level criteria for achieving assay sensitivity. The slope difference criterion is from -13.0 to +17.6% for LC-MS assays with ± 15% bias criteria. When the ratio of endogenous concentration in the tested matrix to the lower limit of quantitation is |
Databáze: | OpenAIRE |