Predictable oxygen concentrations with the bird respirator

Autor: C. H. N. Macdonald
Rok vydání: 1968
Předmět:
Zdroj: Canadian Anaesthetists’ Society Journal. 15:564-578
ISSN: 1496-8975
0008-2856
DOI: 10.1007/bf03004353
Popis: Experimentation with the Bird Mark 7 respirator using the Bird Parallel Inspiratory Fow Mixing Cartridge (999 1289) to provide varying concentrations of oxygen in the inspired gas mixture was carried out. It was found that changes in the absolute pressure did not interfere significantly with the predictability of the oxygen concentrations obtained. However, the oxygen level did vary with changes in flow, being influenced more specifically by the ratio between the flow through the cartridge and the flow through the respirator; i.e., the oxygen concentration was directly proportional to the flow through the cartridge and inversely proportional to the flow through the respirator.