Evaluation of Commercially Available Diagnostic Test Kits for Rubella
Autor: | C. S. Cleghorn, JohnL. Sever, D. L. Madden, G. T. Hazzard, N Tzan, D. V. Vails, Anita C. Ley, Gabriel A. Castellano |
---|---|
Rok vydání: | 1981 |
Předmět: |
business.operation
Antibodies Viral medicine.disease_cause Rubella Hemagglutination tests Diagnostic technology medicine Animals Humans Immunology and Allergy False Positive Reactions False Negative Reactions business.industry Serologic diagnosis Abbott Laboratories Antibody titer virus diseases Diagnostic test Rubella virus Hemagglutination Tests Hemagglutination Inhibition Tests medicine.disease Virology Infectious Diseases Reagent Kits Diagnostic business |
Zdroj: | Journal of Infectious Diseases. 143:578-584 |
ISSN: | 1537-6613 0022-1899 |
DOI: | 10.1093/infdis/143.4.578 |
Popis: | Three standard hemagglutination-inhibition (HAI) methods were compared with 11 commercially available diagnostic test kits for determination of immunity and serologic diagnosis of rubella using a panel of 100 sera. The three standard HAI methods involved removal of serum inhibitors with kaolin, heparin-MnCl2, or dextran sulfate-CaCl2. The HAI kaolin (Flow Laboratories, McLean, Virginia) and Rubelisa (Microbiological Associates Bioproducts, Walkersville, Maryland) kits gave the best specificity as judged by the absence of false-positive results. Rubacell (Abbott Laboratories, Chicago, Illinois), Rubindex (Ortho Diagnostics, Raritan, New Jersey), Fiax (International Diagnostic Technology, Santa Clara, California), and Rubesure (Calbiochem-Behring, La Jolla, California) gave the best sensitivity as judged by the absence of false-negative results. The kits with the highest degree of both specificity and sensitivity were HAI kaolin (Flow), HAI heparin-MnCl2 (Flow), Rubacell (Abbott), and Rubindex (Ortho). In paired sera from five patients with clinical rubella, seroconversions were shown by seven of the kits. One of the kits, Cordia R (Cordis Laboratories, Miami, Florida), showed no significant rise in antibody titer with one pair of sera. Antibody titers in the same serum varied widely between the different kits. |
Databáze: | OpenAIRE |
Externí odkaz: |