Regional left ventricular wall motion from first-pass radionuclide angiocardiography using an Anger camera. Experimental and clinical study
Autor: | Shields Ra, S Walton, D J Rowlands, Testa Hj, C Wrigley, E A Donaldson |
---|---|
Rok vydání: | 1980 |
Předmět: |
medicine.medical_specialty
Heart Diseases business.industry Radiography Heart Ventricles Heart Balloon Imaging phantom law.invention Anger Camera Models Structural law Temporal resolution medicine Cineangiography Humans Radiology Cardiology and Cardiovascular Medicine business Nuclear medicine Radionuclide Imaging Radionuclide angiocardiography Gamma camera Left ventricular wall motion Research Article |
Zdroj: | British heart journal. 44(5) |
ISSN: | 0007-0769 |
Popis: | Experimental studies with a balloon phantom, and clinical studies were performed to evaluate first-pass radionuclide angiocardiography using a gamma camera in the assessment of regional wall motion. The phantom studies showed that choice of isocount contour was not critical in edge detection. Adequate count densities could be achieved, but only at the expense of temporal resolution. The clinical studies disclosed a good correlation with radiography in normal subjects and those with diffuse ventricular disease but a poor correlation in subjects with localised abnormalities of wall motion. |